Pyramid cementless cup

The Pyramid cementless acetabular cup, as well as the revision cup, is manufactured from a titanium alloy and is designed according to the press-fit concept, with a constant interference of 1.7 mm. The threaded hole in the pole is hermetically sealed with a screw after implantation.
Both the Pyramid cementless acetabular cup and the revision cup demonstrate good seating behaviour and excellent primary stability. The unidirectional macrostructure (alternating pyramid-shaped teeth) provides high primary stability with moderate impaction force.
The anchoring structure of both titanium alloy cup types is additionally coated with a two-layer coating of commercially pure titanium and dicalcium phosphate to promote accelerated bone ingrowth and thereby enable safe early mobilization of the patient.
The first layer consists of a 200 µm-thick commercially pure titanium vacuum plasma-sprayed coating (Ti-VPS).
This is overlaid with a 20 µm-thick dicalcium phosphate coating (CaHPO4*2H2O).
Due to the very thin metal shell, the bearing concept has been designed to allow the use of large femoral heads with small cup diameters
(e.g., a 36 mm femoral head fits into a 50 mm cup shell). This provides
a high degree of resistance to dislocation.
Sizes
Standard Cup: Sizes 42-62
High Offset: Sizes 46-70
Accelerated Bone Ingrowth
The very thin, microstructured calcium phosphate layer accelerates bone growth around the implant. This microstructured calcium phosphate layer is no longer present after healing, and the bone is then firmly integrated with the implant. Histological studies have demonstrated that as early as 17 days, a large proportion of the fixation surface is covered with bone.